Amprobe Multitest-1000 Continuity-Tester Manuale Utente Pagina 41

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 122
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 40
main display: 4,15
secondary display: 2,61
3
rd
measurement (not inserted in the medium value)
main display: >1999
secondary display: 2,61
4.5. ANOMALOUS CASES WHICH MAY OCCUR DURING EARTH ρ TESTS
)
- - -
- - - c
If the amperometric circuit is
interrupted, when pressing
START/STOP the instrument can’t read
the minimum current, therefore a
screen similar to the one beside
appears. Make sure that the terminals
are connected correctly and that the
amperometric rod (blue conductor) has
not been driven into non-conductive
ground. If necessary, pour water
around the rod.
rC indicates a
high resistance
value.
SAVE
THE PREVIOUS RESULTS CANNOT BE SAVED
.
)
31 V
VOL
AG
If the instrument measures an
interfering voltage greater than 30V
on the amperometric circuit, it
doesn’t perform the test and a screen
similar to the one beside appears.
Voltage value on the amperometric circuit
SAVE
THE PREVIOUS RESULT CANNOT BE SAVED
.
Vedere la pagina 40
1 2 ... 36 37 38 39 40 41 42 43 44 45 46 ... 121 122

Commenti su questo manuale

Nessun commento